IOQPE03811

Unity Lab Services™ Thermo Scientific™ Liquid Nitrogen Storage Equipment Validation Service

Manufacturer: Unity Lab Services™

Select a Size

Pack Size SKU Availability Price
Each of 1 IOQPE03811-Each-of-1 In Stock ₹ 3,32,828.40

IOQPE03811 - Each of 1

₹ 3,32,828.40

In Stock

Quantity

1

Base Price: ₹ 3,32,828.40

GST (18%): ₹ 59,909.112

Total Price: ₹ 3,92,737.512

For Use With (Equipment)

Thermo Scientific Liquid Nitrogen Storage Equipment including Thermo Scientific CryoExtra™ High-Efficiency Cryogenic Storage Systems Models: CE8180, CE8120, CE8140, CE8194

Description

Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, open door and power loss testing.

Location

Contact your local sales representative for a list of standard and extended coverage areas.

Includes

Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service

Related Products

Img

Unity Lab Services™

IOQPE33913

--

Img

Unity Lab Services™

IOQPE33917

--

Img

Unity Lab Services™

11-675-257

--

Img

Unity Lab Services™

11-675-380

--

Img

Unity Lab Services™

11-675-376

--

Img

Unity Lab Services™

IOQPE03484

--

Img

Unity Lab Services™

11-675-279

--

Img

Unity Lab Services™

IOQPE03493

--

Description

  • IQ Unpacking, Assembly and Installation Verifies that the equipment, manuals, supplies and any other accessories arrived undamaged as specified in the sales order
  • Verifies equipment is assembled and installed and any initial diagnostics tests have been performed
  • Any abnormal event(s) observed during assembly and installation have been documented
  • Utility, Facility and Environment Verifies that the installation site satisfactorily meets manufacturer-specified environmental requirements
  • OQ System Component Information Detailed configuration information for each system component have been recorded
  • OQ Limits: Record a list of manufacturer recommended limits for ensuring that the system is operating as expected
  • Equipment Operational Tests and Results: Test important equipment functions to verify that the equipment operates as intended by the manufacturer and required by the user
  • This is a group of selected important equipment parameters for testing according to the intended use of the equipment
  • Key Tests: Door Ajar, Power Failure, Temperature Alerts

Compare Similar Items

Show Difference

Img

Unity Lab Services™

IOQPE03811

--


For Use With (Equipment):
Thermo Scientific Liquid Nitrogen Storage Equipment including Thermo Scientific CryoExtra™ High-Efficiency Cryogenic Storage Systems Models: CE8180, CE8120, CE8140, CE8194

Description:
Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, open door and power loss testing.

Location:
Contact your local sales representative for a list of standard and extended coverage areas.

Includes:
Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service

Img

Unity Lab Services™

IOQPE03813

--


For Use With (Equipment):
ThermoScientific Laboratory Refrigerator including Thermo Scientific TSX505

Description:
Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, open door and power loss testing.

Location:
Contact your local sales representative for a list of standard and extended coverage areas.

Includes:
Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service

Img

Unity Lab Services™

IOQPE30810

--


For Use With (Equipment):
Thermo Scientific Ultra-Low Temperature Freezers including Ultra-Low Temprature Freezers Models: Thermo Scientific TSX-series (post 2018), Forma 89000-series, HeraFreeze HLE, Revco RLE, Thermo Scientific TLE

Description:
Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, perform empty chamber temperature mapping, open door and power loss testing.

Location:
Contact your local sales representative for a list of standard and extended coverage areas.

Includes:
Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service

Img

Unity Lab Services™

IOQPE33913

--


For Use With (Equipment):
Thermo Scientific Water Purification including Thermo Scientific Barnstead MicroPure, all models

Description:
Validation (IQ/OQ) includes a standardized protocol procedure, IQ: Verify Installation and environmental conditions, OQ: Verify operation to design specification, Flow Adjustment, Water Pressure, Resistivity, Power Failure testing.

Location:
Contact your local sales representative for a list of standard and extended coverage areas.

Includes:
Travel to major metropolitan areas. Does not include Installation. Instrument must be installed, powered-up and set temperature point for 24 hours before on-site service