13-645-587

Thermo Scientific™ Orion™ Versa Star Pro™ pH/ISE/Conductivity/Dissolved Oxygen Multiparameter Benchtop Meter

Manufacturer: Fischer Scientific

Select a Size

Pack Size SKU Availability Price
Each of 1 13-645-587-Each-of-1 In Stock ₹ 4,94,960.00

13-645-587 - Each of 1

₹ 4,94,960.00

In Stock

Quantity

1

Base Price: ₹ 4,94,960.00

GST (18%): ₹ 89,092.80

Total Price: ₹ 5,84,052.80

AC Adapter

Universal 50-60 Hz, 100-240 VAC power adapter (Cat. No. STARA-PWR)

Accuracy (Barometer)

±6 mmHg

Accuracy (ISE)

±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (pH)

±0.002

Accuracy (RDO)

±0.1 mg/L up to 8 mg/L, ±0.2 mg/L from 8 to 20 mg/L, ±10% of reading up to 50 mg/L

Accuracy (Salinity)

0.5% of reading ±1 digit

Advanced ISE Functions

Segmented (point-to-point) slope, non-linear selectable auto-blank, low concentration range stabilit

Backlight Option

Yes

Calibration

pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ

Calibration Points

1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (ISE)

1 to 6

Cell Constant

0.001 to 199.9cm-1

Data Management Details

Transfer single, range or all data points to printer or computer

Description

Orion Versa Star Pro benchtop meter with pH/ISE/mV module, conductivity/TDS/salinity/resistivity module, RDO/DO module and stand

Display Type

Color graphic LCD with user-customizable options

Measurement Modes

pH, mV, Relative mV, ORP, Ion Concentration (ISE) with Temperature, TDS, Conductivity, Salinity

Range (Conductivity)

0.001 μS/cm to 3000 mS/cm

Range (pH)

-2.000 to 20.000

Range (Salinity)

0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution (ISE)

0.0001 minimum, 1 to 4 significant digits (user selectable)

Temperature Range (Metric)

-5.0°C to +105.0°C

Inputs

BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs

Keypad

Numeric with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type

Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with

Methods

10 per channel

Percent Saturation Relative Accuracy (RDO)

±2% saturation ≤ 200% saturation, ±10% saturation > 200% saturation

Percent Saturation Range (RDO)

0.0 to 500.0% saturation

Percent Saturation Resolution (RDO)

0.1, 1% saturation

Power Supply

Universal AC adapter (included)

Pressure Resolution

0.1mmHg

Range (ORP)

±2000.0 mV

Range (RDO)

0.00 to 50.00mg/L

Relative Humidity Range

5 to 85%, non-condensing

Resolution (mV)

0.1mV

Resolution (RDO)

0.01, 0.1mg/L

Resolution (Salinity)

0.01 psu or 0.01 ppt minimum, auto ranging

Salinity Factor

0.0 to 45.0 ppt, automatic correction using conductivity reading or manual entry of sample salinity

Stability

Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy

±0.1°C

Temperature (English) Operating

41°F to 113°F (ambient)

Temperature (Metric) Reference

5°, 10°, 15°, 20°, 25°C

Temperature Range (English)

22.0°F to 221.0° F

Temperature Selection

Manual or automatic with ATC temperature probe

Dimensions (L x W x H)

28 x 24 x 10cm (11 x 9.5 x 3.9 in.), modules add ≤ 2cm (0.8 in.) L & H

Weight (Metric)

1.2 kg, modules add 0.2 kg each

Outputs

USB (2), RS232

Accuracy (Conductivity)

0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS

Accuracy (mV)

±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Polarographic DO)

±0.2 mg/L or ±2% of reading, whichever is greater

Accuracy (Resistivity)

0.5% of reading ±1 digit

Accuracy (TDS)

0.5% of reading ±1 digit

Alarm Output

Set point alarm, high/low limit alarm, calibration due alarm

Barometric Pressure Correction

400 to 850mmHg, automatic using built-in barometer or manual entry

Calibration Modes

Water-saturated air, air-saturated water, manual (Winkler titration), zero point

Calibration Points (Conductivity)

Up to 6

Calibration Points (pH)

1 to 6

Certifications/Compliance

CE, TUV 3-1, FCC Class A

Data Points

2000 with date and time stamp

Channels

4 - customizable with modules

IP Rating

IP 54

Percent Saturation Range Polarographic DO

0.0 to 600.0% saturation

Range (mV)

±2000.0 mV

Range (Resistivity)

2 ohm to 100.0 meg-ohm

Range (TDS)

0.001 ppm to 200.0 ppt

Resolution (pH)

0.1/0.01/0.001

Includes

Versa Star Pro meter, VSTAR-ISE pH/ISE module, VSTAR-CND conductivity module, VSTAR-RD RDO/DO module, electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

Isopotential Point

7.000pH (adjustable), 1.000 ISE (adjustable)

Languages

English, Spanish, French, Italian, German, Chinese, Portuguese and Korean

Memory

Non-volatile memory preserves data log, calibration log and meter settings

Percent Saturation Range (Polarographic DO)

0.0 to 600.0% saturation

Percent Saturation Resolution (Polarographic DO)

0.1, 1% saturation

Percent Saturation Relative Accuracy (Polarographic DO)

±2 % saturation or ±2% of reading whichever is greater

Power Rating

DC input: 9VDC, 1A

Pressure Range

400 to 850mmHg

Range (ISE)

0.0001 to 19990 with selectable units of ppm, Molar, mg/L, %, ppb or none

Range (Polarographic DO)

0.00 to 90.00mg/L

Range (Relative mV)

±2000.0 mV

Resolution (Conductivity)

0.001 μS minimum, auto ranging up to 4 significant digits

Resolution (Polarographic DO)

0.01, 0.1mg/L

Resolution (Resistivity)

1 ohm or 0.1 meg-ohm, auto ranging

Resolution (TDS)

0.001 ppm minimum, auto ranging up to 4 significant digits

Salinity Measurement Modes

Practical salinity (psu) or natural sea water (ppt)

TDS Factor Range

Linear (0.02 to 9.99)

Temperature Compensation

Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off

Temperature (Metric) Operating

5°C to 45°C (ambient)

Temperature Probe Calibration

1 point temperature offset calibration

Temperature Resolution

0.1°C, 0.1°F

Warranty

3 year meter and module warranty

Weight (English)

2.5 lb., modules add 0.5 lb. each

Related Products

Img

Fischer Scientific

13-645-588

--

Img

Fischer Scientific

13-645-569

--

Img

Fischer Scientific

4529615

Accurately measure pH, ion concentr...

Img

Fischer Scientific

4529589

Get premium performance and flexibi...

Img

Fischer Scientific

13-645-571

--

Img

Fischer Scientific

13-645-584

--

Img

Fischer Scientific

13-645-583

--

Img

Fischer Scientific

13-645-545

--

Description

  • Description Get high accuracy, premium performance and ultimate flexibility with the Thermo Scientific™ Orion™ Versa Star Pro™ pH/ISE/Conductivity/Dissolved Oxygen Multiparameter Benchtop Meter
  • Meet your most challenging applications for pH, ion concentration, conductivity, dissolved oxygen and temperature
  • Easily view measurements on the stunning color display and customize the four meter channels with interchangeable modules
  • Log up to 2000 data point sets with time/date stamp and easily transfer logs via USB or RS232 to a printer or computer
  • Select systems with one or two pH/ISE modules
  • Thermo Scientific Orion Versa Star Pro Benchtop Meters are designed to meet your most challenging application with ease
  • Delivering rich functionality and features, you can optimally configure your meter system to meet changing requirements
  • The Orion Versa Star Pro Benchtop Meters offer interchangeable measurement modules that allow multiple users to customize four separate channels to meet their specific requirements
  • Easily view your measurements on a stunning color display with adjustable text sizes
  • Take the guesswork out of meter calibration and setup with onscreen instructions available in a variety of local languages
  • Enjoy quick and simple data entry using the numeric keypad
  • Onscreen measurement stability indicator lets you know exactly when your measurements are ready Selectable read modes include Auto-Read to hold a stable reading, Continuous to show changing readings, Timed to log data at specific intervals and Single-Shot to hold a reading after a specific wait time is reached Audible alarms automatically notify you when a set offset value, high/low limit value and calibration time interval are reached Uniformly mix solutions using up to two meter-controlled stirrer probes that are easy to position and can be quickly rinsed between samples Data log collects and date/time stamps up to 2000 measurement sets with optional electrode ID, sample ID and account ID Calibration log stores 30 most recent calibrations per parameter with viewable graph for multipoint pH and ISE calibrations Ensure measurement integrity by viewing the active calibration log while taking readings Non-volatile meter memory preserves data and settings, even in the event of power loss Meter facilitates data transfers to a computer via USB or RS232 connections Meters include CE, TUV 3-in-1 and FCC Class A certifications 3-year replacement warranty Includes Orion Versa Star pH/ISE Measurement Module: Measure pH, ion concentration, mV, relative mV or ORP with temperature Perform up to a 6 point pH calibration using automatic buffer recognition of NIST, DIN or user-defined buffer groups Perform up to a 6 point ISE calibration with selectable units of ppm, M, mg/L, %, ppb or none Includes Orion Versa Star Conductivity Measurement Module: Measure conductivity, TDS, salinity or resistivity with temperature Perform up to a 6 point conductivity calibration Use linear, non-linear, EP (USP) or off conductivity temperature compensation with 5°C, 10°C, 15°C, 20°C or 25°C reference temperatures Includes Orion Versa Star RDO/DO Measurement Module: Measure dissolved oxygen in percent saturation or mg/L with temperature using Orion RDO optical DO probes or Orion polarographic DO probes Calibrate DO using water-saturated air, air-saturated water, custom value (Winkler titration) and zero point options Automatically compensate DO readings using ATC temperature, manual salinity input and built-in barometer for barometric pressure System Access Feature Orion Versa Star Pro Meters offer a system access feature that is designed to help ensure data integrity by preventing undesired access to various meter functions including date/time setting, data log and calibration log
  • The System Access feature can be easily turned on or off as the needs of your lab change for ultimate flexibility in operation and ease-of-use
  • When enabled, the System Access feature allows assignment of a system administrator and up to 10 user accounts with customizable access to instrument settings, measurement settings, calibrations and logs
  • Use the password protection to secure selected meter functions, while leaving other functions accessible for all users
  • The System Access feature allows you to customize meter access to meet your specific security and data reporting needs
  • Disclaimer to internal research use restriction: Any restrictions on the purchaser to utilize the product(s) for internal research purposes only does not apply to the product(s) on the web pages associated with this statement.

Compare Similar Items

Show Difference

Img

Fischer Scientific

13-645-587

--


AC Adapter:
Universal 50-60 Hz, 100-240 VAC power adapter (Cat. No. STARA-PWR)

Accuracy (Barometer):
±6 mmHg

Accuracy (ISE):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (pH):
±0.002

Accuracy (RDO):
±0.1 mg/L up to 8 mg/L, ±0.2 mg/L from 8 to 20 mg/L, ±10% of reading up to 50 mg/L

Accuracy (Salinity):
0.5% of reading ±1 digit

Advanced ISE Functions:
Segmented (point-to-point) slope, non-linear selectable auto-blank, low concentration range stabilit

Backlight Option:
Yes

Calibration:
pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ

Calibration Points:
1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (ISE):
1 to 6

Cell Constant:
0.001 to 199.9cm-1

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Versa Star Pro benchtop meter with pH/ISE/mV module, conductivity/TDS/salinity/resistivity module, RDO/DO module and stand

Display Type:
Color graphic LCD with user-customizable options

Measurement Modes:
pH, mV, Relative mV, ORP, Ion Concentration (ISE) with Temperature, TDS, Conductivity, Salinity

Range (Conductivity):
0.001 μS/cm to 3000 mS/cm

Range (pH):
-2.000 to 20.000

Range (Salinity):
0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution (ISE):
0.0001 minimum, 1 to 4 significant digits (user selectable)

Temperature Range (Metric):
-5.0°C to +105.0°C

Inputs:
BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs

Keypad:
Numeric with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with

Methods:
10 per channel

Percent Saturation Relative Accuracy (RDO):
±2% saturation ≤ 200% saturation, ±10% saturation > 200% saturation

Percent Saturation Range (RDO):
0.0 to 500.0% saturation

Percent Saturation Resolution (RDO):
0.1, 1% saturation

Power Supply:
Universal AC adapter (included)

Pressure Resolution:
0.1mmHg

Range (ORP):
±2000.0 mV

Range (RDO):
0.00 to 50.00mg/L

Relative Humidity Range:
5 to 85%, non-condensing

Resolution (mV):
0.1mV

Resolution (RDO):
0.01, 0.1mg/L

Resolution (Salinity):
0.01 psu or 0.01 ppt minimum, auto ranging

Salinity Factor:
0.0 to 45.0 ppt, automatic correction using conductivity reading or manual entry of sample salinity

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
5°, 10°, 15°, 20°, 25°C

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
28 x 24 x 10cm (11 x 9.5 x 3.9 in.), modules add ≤ 2cm (0.8 in.) L & H

Weight (Metric):
1.2 kg, modules add 0.2 kg each

Outputs:
USB (2), RS232

Accuracy (Conductivity):
0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Polarographic DO):
±0.2 mg/L or ±2% of reading, whichever is greater

Accuracy (Resistivity):
0.5% of reading ±1 digit

Accuracy (TDS):
0.5% of reading ±1 digit

Alarm Output:
Set point alarm, high/low limit alarm, calibration due alarm

Barometric Pressure Correction:
400 to 850mmHg, automatic using built-in barometer or manual entry

Calibration Modes:
Water-saturated air, air-saturated water, manual (Winkler titration), zero point

Calibration Points (Conductivity):
Up to 6

Calibration Points (pH):
1 to 6

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
2000 with date and time stamp

Channels:
4 - customizable with modules

IP Rating:
IP 54

Percent Saturation Range Polarographic DO:
0.0 to 600.0% saturation

Range (mV):
±2000.0 mV

Range (Resistivity):
2 ohm to 100.0 meg-ohm

Range (TDS):
0.001 ppm to 200.0 ppt

Resolution (pH):
0.1/0.01/0.001

Includes:
Versa Star Pro meter, VSTAR-ISE pH/ISE module, VSTAR-CND conductivity module, VSTAR-RD RDO/DO module, electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

Isopotential Point:
7.000pH (adjustable), 1.000 ISE (adjustable)

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Percent Saturation Range (Polarographic DO):
0.0 to 600.0% saturation

Percent Saturation Resolution (Polarographic DO):
0.1, 1% saturation

Percent Saturation Relative Accuracy (Polarographic DO):
±2 % saturation or ±2% of reading whichever is greater

Power Rating:
DC input: 9VDC, 1A

Pressure Range:
400 to 850mmHg

Range (ISE):
0.0001 to 19990 with selectable units of ppm, Molar, mg/L, %, ppb or none

Range (Polarographic DO):
0.00 to 90.00mg/L

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
0.001 μS minimum, auto ranging up to 4 significant digits

Resolution (Polarographic DO):
0.01, 0.1mg/L

Resolution (Resistivity):
1 ohm or 0.1 meg-ohm, auto ranging

Resolution (TDS):
0.001 ppm minimum, auto ranging up to 4 significant digits

Salinity Measurement Modes:
Practical salinity (psu) or natural sea water (ppt)

TDS Factor Range:
Linear (0.02 to 9.99)

Temperature Compensation:
Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 year meter and module warranty

Weight (English):
2.5 lb., modules add 0.5 lb. each

Img

Fischer Scientific

13-645-588

--


AC Adapter:
Universal 50-60 Hz, 100-240 VAC power adapter (Cat. No. STARA-PWR)

Accuracy (Barometer):
±6 mmHg

Accuracy (ISE):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (pH):
±0.002

Accuracy (RDO):
±0.1 mg/L up to 8 mg/L, ±0.2 mg/L from 8 to 20 mg/L, ±10% of reading up to 50 mg/L

Accuracy (Salinity):
0.5% of reading ±1 digit

Advanced ISE Functions:
Timed endpoint, linear point to point, non-linear selectable auto-blank, low concentration range sta

Backlight Option:
Yes

Calibration:
pH with calibration editing option, ion concentration (ISE) with calibration editing option, relativ

Calibration Points:
1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (ISE):
1 to 6

Cell Constant:
0.001 to 199.9cm-1

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Versa Star Pro Benchtop Meter with VSTAR-ISE pH/mV/ISE/Temperature Module; VSTAR-CND Conductivity/TDS/Resistivity/Salinity/Temperature Module; VSTAR-RD RDO/DO/Temperature Module Kit

Display Type:
Color graphic LCD with user-customizable options

Measurement Modes:
pH, mV, Relative mV, ORP, Ion Concentration (ISE) with Temperature, TDS, Conductivity, Salinity

Range (Conductivity):
0.001 μS/cm to 3000 mS/cm

Range (pH):
-2.000 to 20.000

Range (Salinity):
0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution (ISE):
0.0001 minimum, 1 to 4 significant digits (user selectable)

Temperature Range (Metric):
-5.0°C to +105.0°C

Inputs:
BNC (pH or ISE), 8-pin Mini-DIN (temperature), reference pin-tip, 8-pin Mini-DIN (conductivity with temperature probe), 9-pin Mini-DIN (RDO or polarographic DO with temperature probe), 2 stirrer inputs

Keypad:
Numeric with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read, Timed and Single Shot measure modes; manual data logging with

Methods:
10 per channel with password protection

Percent Saturation Relative Accuracy (RDO):
±2% saturation ≤ 200% saturation, ±10% saturation > 200% saturation

Percent Saturation Range (RDO):
0.0 to 500.0% saturation

Percent Saturation Resolution (RDO):
0.1, 1% saturation

Power Supply:
Universal AC adapter (included)

Pressure Resolution:
0.1mmHg

Range (ORP):
±2000.0 mV

Range (RDO):
0.00 to 50.00mg/L

Relative Humidity Range:
5 to 85%, non-condensing

Resolution (mV):
0.1mV

Resolution (RDO):
0.01, 0.1mg/L

Resolution (Salinity):
0.01 psu or 0.01 ppt minimum, auto ranging

Salinity Factor:
0.0 to 45.0 ppt, automatic correction using conductivity reading or manual entry of sample salinity

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
5°, 10°, 15°, 20°, 25°C

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
28 x 24 x 10cm (11 x 9.5 x 3.9 in.), modules add ≤ 2cm (0.8 in.) L & H

Weight (Metric):
1.2 kg, modules add 0.2 kg each

Outputs:
USB (2), RS232

Accuracy (Conductivity):
0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Polarographic DO):
±0.2 mg/L or ±2% of reading, whichever is greater

Accuracy (Resistivity):
0.5% of reading ±1 digit

Accuracy (TDS):
0.5% of reading ±1 digit

Alarm Output:
Set point alarm, high/low limit alarm, calibration due alarm

Barometric Pressure Correction:
400 to 850mmHg, automatic using built-in barometer or manual entry

Calibration Modes:
Water-saturated air, air-saturated water, manual (Winkler titration), zero point

Calibration Points (Conductivity):
Up to 6

Calibration Points (pH):
1 to 6

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
2000 with date and time stamp

Channels:
4 - customizable with modules

IP Rating:
IP 54

Percent Saturation Range Polarographic DO:
0.0 to 600.0% saturation

Range (mV):
±2000.0 mV

Range (Resistivity):
2 ohm to 100.0 meg-ohm

Range (TDS):
0.001 ppm to 200.0 ppt

Resolution (pH):
0.1/0.01/0.001

Includes:
Versa Star Pro meter, VSTAR-ISE pH/ISE module, VSTAR-CND conductivity module, VSTAR-RD RDO/DO module, 8157BNUMD ROSS Ultra Triode epoxy-body pH/ATC electrode, 013005MD Orion DuraProbe 4-cell (K=0.475) conductivity sensor, 083005MD Orion polarographic dissolved oxygen sensor, 096019 Orion Star stirrer probe, 810199 ROSS pH buffer and solution kit, 011007 Orion 1413µS conductivity standard, 080513 DO maintenance kit, BOD adapter/funnel/stirrer, electrode stand, universal power adapter, literature CD, printed quick start guide, computer interface cable and meter test certificate

Isopotential Point:
7.000pH (adjustable), 1.000 ISE (adjustable)

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Percent Saturation Range (Polarographic DO):
0.0 to 600.0% saturation

Percent Saturation Resolution (Polarographic DO):
0.1, 1% saturation

Percent Saturation Relative Accuracy (Polarographic DO):
±2 % saturation or ±2% of reading whichever is greater

Power Rating:
DC input: 9VDC, 1A

Pressure Range:
400 to 850mmHg

Range (ISE):
0.0001 to 19990 with selectable units of ppm, Molar, mg/L, %, ppb or none

Range (Polarographic DO):
0.00 to 90.00mg/L

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
0.001 μS minimum, auto ranging up to 4 significant digits

Resolution (Polarographic DO):
0.01, 0.1mg/L

Resolution (Resistivity):
1 ohm or 0.1 meg-ohm, auto ranging

Resolution (TDS):
0.001 ppm minimum, auto ranging up to 4 significant digits

Salinity Measurement Modes:
Practical salinity (psu) or natural sea water (ppt)

TDS Factor Range:
Linear (0.02 to 9.99)

Temperature Compensation:
Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 year meter and module warranty

Weight (English):
2.5 lb., modules add 0.5 lb. each

Img

Fischer Scientific

13-650-110

--


AC Adapter:
__

Accuracy (Barometer):
__

Accuracy (ISE):
__

Accuracy (pH):
__

Accuracy (RDO):
__

Accuracy (Salinity):
__

Advanced ISE Functions:
__

Backlight Option:
__

Calibration:
__

Calibration Points:
__

Calibration Points (ISE):
__

Cell Constant:
__

Data Management Details:
__

Description:
Silicon Oil and Cloth Kit

Display Type:
__

Measurement Modes:
__

Range (Conductivity):
__

Range (pH):
__

Range (Salinity):
__

Resolution (ISE):
__

Temperature Range (Metric):
__

Inputs:
__

Keypad:
__

Log Function Type:
__

Methods:
__

Percent Saturation Relative Accuracy (RDO):
__

Percent Saturation Range (RDO):
__

Percent Saturation Resolution (RDO):
__

Power Supply:
__

Pressure Resolution:
__

Range (ORP):
__

Range (RDO):
__

Relative Humidity Range:
__

Resolution (mV):
__

Resolution (RDO):
__

Resolution (Salinity):
__

Salinity Factor:
__

Stability:
__

Temperature Accuracy:
__

Temperature (English) Operating:
__

Temperature (Metric) Reference:
__

Temperature Range (English):
__

Temperature Selection:
__

Dimensions (L x W x H):
__

Weight (Metric):
__

Outputs:
__

Accuracy (Conductivity):
__

Accuracy (mV):
__

Accuracy (Polarographic DO):
__

Accuracy (Resistivity):
__

Accuracy (TDS):
__

Alarm Output:
__

Barometric Pressure Correction:
__

Calibration Modes:
__

Calibration Points (Conductivity):
__

Calibration Points (pH):
__

Certifications/Compliance:
__

Data Points:
__

Channels:
__

IP Rating:
__

Percent Saturation Range Polarographic DO:
__

Range (mV):
__

Range (Resistivity):
__

Range (TDS):
__

Resolution (pH):
__

Includes:
__

Isopotential Point:
__

Memory:
__

Percent Saturation Range (Polarographic DO):
__

Percent Saturation Resolution (Polarographic DO):
__

Percent Saturation Relative Accuracy (Polarographic DO):
__

Power Rating:
__

Pressure Range:
__

Range (ISE):
__

Range (Polarographic DO):
__

Range (Relative mV):
__

Resolution (Conductivity):
__

Resolution (Polarographic DO):
__

Resolution (Resistivity):
__

Resolution (TDS):
__

Salinity Measurement Modes:
__

TDS Factor Range:
__

Temperature Compensation:
__

Temperature (Metric) Operating:
__

Temperature Probe Calibration:
__

Temperature Resolution:
__

Warranty:
__

Weight (English):
__

Img

Fischer Scientific

13-650-116

--


AC Adapter:
__

Accuracy (Barometer):
__

Accuracy (ISE):
__

Accuracy (pH):
__

Accuracy (RDO):
__

Accuracy (Salinity):
__

Advanced ISE Functions:
__

Backlight Option:
__

Calibration:
__

Calibration Points:
__

Calibration Points (ISE):
__

Cell Constant:
__

Data Management Details:
__

Description:
AQUAfast ammonia LR Auto-ID ampoules, 30 pack

Display Type:
__

Measurement Modes:
__

Range (Conductivity):
__

Range (pH):
__

Range (Salinity):
__

Resolution (ISE):
__

Temperature Range (Metric):
__

Inputs:
__

Keypad:
__

Log Function Type:
__

Methods:
__

Percent Saturation Relative Accuracy (RDO):
__

Percent Saturation Range (RDO):
__

Percent Saturation Resolution (RDO):
__

Power Supply:
__

Pressure Resolution:
__

Range (ORP):
__

Range (RDO):
__

Relative Humidity Range:
__

Resolution (mV):
__

Resolution (RDO):
__

Resolution (Salinity):
__

Salinity Factor:
__

Stability:
__

Temperature Accuracy:
__

Temperature (English) Operating:
__

Temperature (Metric) Reference:
__

Temperature Range (English):
__

Temperature Selection:
__

Dimensions (L x W x H):
__

Weight (Metric):
__

Outputs:
__

Accuracy (Conductivity):
__

Accuracy (mV):
__

Accuracy (Polarographic DO):
__

Accuracy (Resistivity):
__

Accuracy (TDS):
__

Alarm Output:
__

Barometric Pressure Correction:
__

Calibration Modes:
__

Calibration Points (Conductivity):
__

Calibration Points (pH):
__

Certifications/Compliance:
__

Data Points:
__

Channels:
__

IP Rating:
__

Percent Saturation Range Polarographic DO:
__

Range (mV):
__

Range (Resistivity):
__

Range (TDS):
__

Resolution (pH):
__

Includes:
30 Auto-ID Ampoules

Isopotential Point:
__

Memory:
__

Percent Saturation Range (Polarographic DO):
__

Percent Saturation Resolution (Polarographic DO):
__

Percent Saturation Relative Accuracy (Polarographic DO):
__

Power Rating:
__

Pressure Range:
__

Range (ISE):
__

Range (Polarographic DO):
__

Range (Relative mV):
__

Resolution (Conductivity):
__

Resolution (Polarographic DO):
__

Resolution (Resistivity):
__

Resolution (TDS):
__

Salinity Measurement Modes:
__

TDS Factor Range:
__

Temperature Compensation:
__

Temperature (Metric) Operating:
__

Temperature Probe Calibration:
__

Temperature Resolution:
__

Warranty:
__

Weight (English):
__