S72305

GENESYS 140 Vis Spectrophotometer

Manufacturer: Thermo Scientific

The price for this product is unavailable. Please request a quote

Baseline Flatness

±0.005 A

Connections

Single USB-ADuplex USB-AExport data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, availablePrint via USB, Ethernet or Wi-Fi USB adaptor available

Detector Type

Dual Silicon Photodiodes

Dimensions (L x W x H)

35.5 x 38.5 x 19.5 cm

Spectral Bandwidth

5 nm

Drift

(At 500 nm after 1 hour warmup) <0.0010 A/hr.

Keypad

Touchscreen

Noise

(RMS at 500nm 60 consecutive measurements)≤0.00020A at 0A at 260 and 500nm≤0.00030A at 1A at 260 and 500nm≤0.00040A at 2A at 260 and 500nm

Photometric Accuracy

±0.002 A at 0.5 A, ±0.004A at 1.0 A, ±0.008 A at 2.0 A

Photometric Repeatability

±0.001 A at 1 A measured at 1.0 A at 546 nm

Scan Speed

Automatic – up to 1800 nm/min.

Warranty

2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase

Wavelength Data Interval

0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Weight (Metric)

7.5 kg

Compartments

Accommodates (optional):8-position cell changer4-position cell changer (long path cells)Peltier thermostatted cell holder (20-60 °C)Sipper accessoryFiber optic probe coupler

Description

GENESYS 140 Visible Spectrophotometer

Display

7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels

Light Source

Tungsten Halogen

Wavelength Range

325 to 1100 nm

Electrical Requirements

External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

Lamp Life

>1000 hrs.

Optical Design

Double Beam

Photometric Range

−3A to +3.5A

Printer Included

Snap-on printer available

Stray Light

<0.05%T at 340 nm and 400 nm

Wavelength Accuracy

±0.5 nm

Wavelength Repeatability

< ±0.2 nm

Related Products

Img

Thermo Scientific

S72306

--

Img

Thermo Scientific™

14-385-355

--

Img

Thermo Scientific

S72307

--

Img

Thermo Scientific™

14-385-354

--

Img

Thermo Scientific

S72303

--

Img

Thermo Scientific™

14-385-355PM

--

Img

Thermo Scientific

S72304

--

Img

Thermo Scientific

S28454

--

Description

  • The GENESYS 150 UV-Vis Spectrophotometer provides automation for high-throughput options and room-light resistance, allowing lid-open operation for improved speed and convenience
  • A full line of accessories, including automated cell changers and cell holders, sipper, and fiber optic probes are designed to simplify sampling and accommodate high throughput measurements and temperature control
  • The GENESYS 140 Vis Spectrophotometer is a visible spectral region version in the instrumental class
  • Both instruments feature data resolution from 0.2 nm to 5 nm
  • Features: 7-inch high-resolution touchscreen user interface Onboard methods, reporting and data export Networking capable (Wi-Fi ready) Room light resistance for “open-lid” operation Sipper accessory Removable, washable sample compartment liner with magnetic placement and hold-down Accommodates cells up to 100 mm pathlength (Optional) 8-position cell changer with switchable turret (Optional) 4-position cell changer (long path cells) (Optional) Peltier-thermostatted single cell holder (20 °C - 60 °C) (Optional) Accessory for test-tubes up to 25 mm diameter and 150 mm height Designed for: Advanced teaching labs R&D departments Applications requiring high throughput or temperature control

Compare Similar Items

Show Difference

Img

Thermo Scientific

S72305

--


Baseline Flatness:
±0.005 A

Connections:
Single USB-ADuplex USB-AExport data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, availablePrint via USB, Ethernet or Wi-Fi USB adaptor available

Detector Type:
Dual Silicon Photodiodes

Dimensions (L x W x H):
35.5 x 38.5 x 19.5 cm

Spectral Bandwidth:
5 nm

Drift:
(At 500 nm after 1 hour warmup) <0.0010 A/hr.

Keypad:
Touchscreen

Noise:
(RMS at 500nm 60 consecutive measurements)≤0.00020A at 0A at 260 and 500nm≤0.00030A at 1A at 260 and 500nm≤0.00040A at 2A at 260 and 500nm

Photometric Accuracy:
±0.002 A at 0.5 A, ±0.004A at 1.0 A, ±0.008 A at 2.0 A

Photometric Repeatability:
±0.001 A at 1 A measured at 1.0 A at 546 nm

Scan Speed:
Automatic – up to 1800 nm/min.

Warranty:
2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase

Wavelength Data Interval:
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Weight (Metric):
7.5 kg

Compartments:
Accommodates (optional):8-position cell changer4-position cell changer (long path cells)Peltier thermostatted cell holder (20-60 °C)Sipper accessoryFiber optic probe coupler

Description:
GENESYS 140 Visible Spectrophotometer

Display:
7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels

Light Source:
Tungsten Halogen

Wavelength Range:
325 to 1100 nm

Electrical Requirements:
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

Lamp Life:
>1000 hrs.

Optical Design:
Double Beam

Photometric Range:
−3A to +3.5A

Printer Included:
Snap-on printer available

Stray Light:
<0.05%T at 340 nm and 400 nm

Wavelength Accuracy:
±0.5 nm

Wavelength Repeatability:
< ±0.2 nm

Img

Thermo Scientific

S72306

--


Baseline Flatness:
±0.002 A

Connections:
Single USB-ADuplex USB-AExport data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, availablePrint via USB, Ethernet or Wi-Fi USB adaptor available

Detector Type:
Dual Silicon Photodiodes

Dimensions (L x W x H):
35.5 x 38.5 x 19.5 cm

Spectral Bandwidth:
2 nm

Drift:
(At 500 nm after 1 hour warmup) <0.0005 A/hr.

Keypad:
Touchscreen

Noise:
(RMS at 500nm 60 consecutive measurements)≤0.00020A at 0A at 260 and 500nm≤0.00030A at 1A at 260 and 500nm≤0.00040A at 2A at 260 and 500nm

Photometric Accuracy:
±0.002 A at 0.5 A, ±0.004A at 1.0A, ±0.008 A at 2.0 A

Photometric Repeatability:
±0.001 A at 1 A measured at 1.0 A at 546 nm

Scan Speed:
Slow, medium and fast - up to 1600 nm/min.

Warranty:
2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase

Wavelength Data Interval:
0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Weight (Metric):
7.5 kg

Compartments:
Accommodates (optional):8-position cell changer4-position cell changer (long path cells)Peltier thermostatted cell holder (20-60 °C)Sipper accessoryFiber optic probe coupler

Description:
GENESYS 150 UV-Visible Spectrophotometer

Display:
7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels

Light Source:
__

Wavelength Range:
190 to 1100 nm

Electrical Requirements:
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

Lamp Life:
> 5 years typical, 3 years guaranteed

Optical Design:
Double Beam

Photometric Range:
–2A to +3.5A

Printer Included:
Snap-on printer available

Stray Light:
< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2)

Wavelength Accuracy:
±0.5 nm

Wavelength Repeatability:
< ±0.2 nm

Img

Thermo Scientific

S72307

--


Baseline Flatness:
±0.002 A

Connections:
Single USB-ADuplex USB-AExport data to network or PC via USB, Ethernet, or Wi-Fi USB adaptor, availablePrint via USB, Ethernet or Wi-Fi USB adaptor available

Detector Type:
Dual Silicon Photodiodes

Dimensions (L x W x H):
35.5 x 38.5 x 19.5 cm

Spectral Bandwidth:
2 nm

Drift:
(At 500 nm after 1 hour warmup) <0.0005 A/hr.

Keypad:
Touchscreen

Noise:
(RMS at 500nm 60 consecutive measurements)≤0.00020A at 0A at 260 and 500nm≤0.00030A at 1A at 260 and 500nm≤0.00040A at 2A at 260 and 500nm

Photometric Accuracy:
±0.002 A at 0.5 A, ±0.004A at 1.0A, ±0.008 A at 2.0 A

Photometric Repeatability:
±0.001 A at 1 A measured at 1.0 A at 546 nm

Scan Speed:
Slow, medium and fast - up to 1600 nm/min.

Warranty:
2 years standard with 1 year extension upon registration of the instrument with Thermo Fisher Scientific within 6 months of purchase

Wavelength Data Interval:
0.1 nm, 0.2 nm, 0.5 nm, 1 nm, 2 nm, 5 nm

Weight (Metric):
7.5 kg

Compartments:
8-position cell changer (standard)Also accommodates (optional):4-position cell changer (long path cells)Peltier thermostatted cell holder (20-60 °C)Sipper accessoryFiber optic probe coupler

Description:
GENESYS 180 UV-Vis Spectrophotometer

Display:
7 in. High Definition Color Touchscreen, Tiltable, 800 x 1280 pixels

Light Source:
__

Wavelength Range:
190 to 1100 nm

Electrical Requirements:
External AC to DC converter. Voltage and Frequency (Hz) selected automatically, 100/240 V, 50/60 Hz

Lamp Life:
> 5 years typical, 3 years guaranteed

Optical Design:
Double Beam

Photometric Range:
–2A to +3.5A

Printer Included:
Snap-on printer available

Stray Light:
< 1.0%T 198 nm (KCl) , <0.05%T at 220 nm (NaI), <0.03%T at 340 nm (NaNO2)

Wavelength Accuracy:
±0.5 nm

Wavelength Repeatability:
< ±0.2 nm

Img

Bel-Art™ SP Scienceware™

S72309

--


Baseline Flatness:
__

Connections:
__

Detector Type:
__

Dimensions (L x W x H):
__

Spectral Bandwidth:
__

Drift:
__

Keypad:
__

Noise:
__

Photometric Accuracy:
__

Photometric Repeatability:
__

Scan Speed:
__

Warranty:
__

Wavelength Data Interval:
__

Weight (Metric):
__

Compartments:
__

Description:
__

Display:
__

Light Source:
__

Wavelength Range:
__

Electrical Requirements:
__

Lamp Life:
__

Optical Design:
__

Photometric Range:
__

Printer Included:
__

Stray Light:
__

Wavelength Accuracy:
__

Wavelength Repeatability:
__