13-645-547PM

Thermo Scientific™ Orion™ Star A215 pH/conductivity Benchtop Multiparameter Meter Kit PROMO

Manufacturer: Thermo Scientific™

The price for this product is unavailable. Please request a quote

AC Adapter

Universal 50 to 60 Hz, 100-240 VAC power adapter (Cat. No. 1010003)

Accuracy (Conductivity)

0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS

Accuracy (pH)

±0.002

Accuracy (Salinity)

0.5% of reading ±1 digit

Alarm Output

High/low limit alarm, calibration due alarm

Battery Life

800 hr.

Calibration

pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing op

Calibration Points (Conductivity)

Up to 5

Cell Constant

0.001 to 199.9 cm-1

Data Management Details

Transfer single, range or all data points to printer or computer

Description

Orion Star A215 pH/conductivity benchtop meter kit with ROSS Ultra Triode epoxy-body pH/ATC electrode, DuraProbe 4-cell conductivity sensor, solutions and stand.

Display Type

LCD Graphic, Backlit

Measurement Modes

pH, mV, Relative mV, TDS, ORP with Temperature, Salinity, Conductivity, Resistivity with Temperature

Range (mV)

±2000.0 mV

Range (Resistivity)

2 ohm to 100.0 meg-ohm

Range (TDS)

0.001 to 200.0 ppm

Temperature Range (Metric)

-5.0°C to +105.0°C

Isopotential Point

7 pH

Languages

English, Spanish, French, Italian, German and Chinese; Portuguese and Korean available with software update

Memory

Non-volatile memory preserves data log, calibration log and meter settings

Model

Orion Star A215 pH/mV/RmV/ORP/Conductivity/TDS/Salinity/Resistivity/Temperature Benchtop Meter

Range (ORP)

±2000.0 mV

Relative Humidity Range

5 to 85 %, non-condensing

Resolution (mV)

0.1 mV

Resolution (Salinity)

0.01 psu or 0.01 ppt minimum, auto ranging

Salinity Measurement Modes

Practical salinity (psu) or natural sea water (ppt)

TDS Factor Range

Linear (0.02 to 9.99)

Temperature Compensation

Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off

Temperature (Metric) Operating

5°C to 45°C (ambient)

Temperature Probe Calibration

1 point temperature offset calibration

Temperature Resolution

0.1°C, 0.1°F

Warranty

3 Year Meter

Weight (English)

2 lb

Outputs

USB, RS232

Accuracy (mV)

±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Resistivity)

0.5% of reading ±1 digit

Accuracy (TDS)

0.5% of reading ±1 digit

Backlight Option

Yes, selectable

Battery Type

4 AA (sold separately)

Calibration Points

1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (pH)

1 to 5

Certifications/Compliance

CE, TUV 3-1, FCC Class A

Data Points

2000 with date and time stamp

Channels

2

IP Rating

IP 54

Range (Conductivity)

0.001 μS/cm to 3000 mS/cm

Range (pH)

-2.000 to 20.000

Range (Salinity)

0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution (pH)

0.1/0.01/0.001

Inputs

BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)

Keypad

Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type

Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m

Methods

10 per channel with password protection

Power Supply

Universal AC adapter (included) or 4 AA batteries (sold separately)

Range (Relative mV)

±2000.0 mV

Resolution (Conductivity)

0.001 μS minimum, auto ranging up to 4 significant digits

Resolution (Resistivity)

1 ohm or 0.1 meg-ohm, auto ranging

Resolution (TDS)

0.001 ppm minimum, auto ranging up to 4 significant digits

Stability

Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy

±0.1°C

Temperature (English) Operating

41°F to 113°F (ambient)

Temperature (Metric) Reference

5, 10, 15, 20, 25°C

Temperature Range (English)

22.0°F to 221.0° F

Temperature Selection

Manual or automatic with ATC temperature probe

Dimensions (L x W x H)

24 x 18 x 11 cm (9.5 x 7.1 x 4.3 in.)

Weight (Metric)

0.9 kg

Related Products

Img

Fischer Scientific

13-645-545

--

Img

Fischer Scientific

4529612

Accurately measure pH, conductivity...

Img

Fischer Scientific

4529602

Reliably measure pH, conductivity a...

Img

Fischer Scientific

13-645-563

--

Img

Fischer Scientific

13-645-547

--

Img

Thermo Scientific™

13-645-565PM

--

Img

Fischer Scientific

13-645-565

--

Img

Fischer Scientific

4529592

Get premium performance and flexibi...

Description

  • The Thermo Scientific Orion Star A215 pH/ Conductivity Benchtop Meter is the ideal choice for any lab
  • Get the information you need quickly and easily from the large, backlit graphic LCD display
  • Informative display shows individual pH/mV/RmV or conductivity/TDS/salinity/ resistivity results with temperature, or all results at once
  • Additional helpful information includes electrode status, time, date, sample ID, user ID and calibration
  • For advanced users, features such as stability and averaging choices provide additional options
  • Anyone can use this meter with its plain language screen prompts, soft keys that update for easy selection and a multilanguage interface that supports English, Spanish, French, Italian, German and Chinese Don't miss a reading — AUTO-READ™ locks in the stable reading on your screen, ready indicator alerts when readings are stable and timed reading gathers data in specific time intervals Up to five point pH calibration with automatic recognition for USA/NIST and DIN buffers and calibration editing to fix errors without a complete recalibration Selectable cell constant allows for use with 2- or 4-cell conductivity cells For accurate conductivity results, selectable reading reference temperatures of 5°, 10°, 15°, 20° or 25°C with linear, nonlinear, nLFu or EP curve options Choice of practical salinity or natural sea water curves for salinity readings Linear or ISO/EN 27888 curves for TDS Nonvolatile memory holds up to 2000 data points with time and date stamp Easily transfer data and keep meter software up-to-date with the USB and RS232 ports and complimentary data analysis software Mix samples without a stir plate with direct control of the Orion Star stirrer probe (096019, sold separately) Included electrode arm and newly designed probe holder make it easier to maintain and place probes into samples Works with almost every AC power source with the included universal power adapter or with four AA batteries (sold separately) to run the meter on DC power IP54-rated housing handles splashes and is wall-mountable if table space is a problem Solutions included: 14/13 conductivity standard, pH buffer and solution kit (10,7,4), storage solution and cleaning solution

Compare Similar Items

Show Difference

Img

Thermo Scientific™

13-645-547PM

--


AC Adapter:
Universal 50 to 60 Hz, 100-240 VAC power adapter (Cat. No. 1010003)

Accuracy (Conductivity):
0.5% of reading ±1 digit > 3 μS; 0.5% of reading ±0.01 μS ≤ 3 μS

Accuracy (pH):
±0.002

Accuracy (Salinity):
0.5% of reading ±1 digit

Alarm Output:
High/low limit alarm, calibration due alarm

Battery Life:
800 hr.

Calibration:
pH with calibration editing option, relative mV (RmV), ORP, conductivity with calibration editing op

Calibration Points (Conductivity):
Up to 5

Cell Constant:
0.001 to 199.9 cm-1

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Star A215 pH/conductivity benchtop meter kit with ROSS Ultra Triode epoxy-body pH/ATC electrode, DuraProbe 4-cell conductivity sensor, solutions and stand.

Display Type:
LCD Graphic, Backlit

Measurement Modes:
pH, mV, Relative mV, TDS, ORP with Temperature, Salinity, Conductivity, Resistivity with Temperature

Range (mV):
±2000.0 mV

Range (Resistivity):
2 ohm to 100.0 meg-ohm

Range (TDS):
0.001 to 200.0 ppm

Temperature Range (Metric):
-5.0°C to +105.0°C

Isopotential Point:
7 pH

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Model:
Orion Star A215 pH/mV/RmV/ORP/Conductivity/TDS/Salinity/Resistivity/Temperature Benchtop Meter

Range (ORP):
±2000.0 mV

Relative Humidity Range:
5 to 85 %, non-condensing

Resolution (mV):
0.1 mV

Resolution (Salinity):
0.01 psu or 0.01 ppt minimum, auto ranging

Salinity Measurement Modes:
Practical salinity (psu) or natural sea water (ppt)

TDS Factor Range:
Linear (0.02 to 9.99)

Temperature Compensation:
Linear, nLFn (non-linear ultra pure water), nLFu (non-linear ultra pure degassed water), EP (off with warning) or off

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 Year Meter

Weight (English):
2 lb

Outputs:
USB, RS232

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Resistivity):
0.5% of reading ±1 digit

Accuracy (TDS):
0.5% of reading ±1 digit

Backlight Option:
Yes, selectable

Battery Type:
4 AA (sold separately)

Calibration Points:
1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (pH):
1 to 5

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
2000 with date and time stamp

Channels:
2

IP Rating:
IP 54

Range (Conductivity):
0.001 μS/cm to 3000 mS/cm

Range (pH):
-2.000 to 20.000

Range (Salinity):
0.06 to 80.00 psu, 0.05 to 42.00 ppt

Resolution (pH):
0.1/0.01/0.001

Inputs:
BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (conductivity sensor with built-in temperature or ATC temperature probe), stirrer input (stirrer probe)

Keypad:
Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m

Methods:
10 per channel with password protection

Power Supply:
Universal AC adapter (included) or 4 AA batteries (sold separately)

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
0.001 μS minimum, auto ranging up to 4 significant digits

Resolution (Resistivity):
1 ohm or 0.1 meg-ohm, auto ranging

Resolution (TDS):
0.001 ppm minimum, auto ranging up to 4 significant digits

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
5, 10, 15, 20, 25°C

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
24 x 18 x 11 cm (9.5 x 7.1 x 4.3 in.)

Weight (Metric):
0.9 kg

Img

Fischer Scientific

13-645-548

--


AC Adapter:
Universal 50-60 Hz, 100-240 VAC power adapter (Cat. No. 1010003)

Accuracy (Conductivity):
__

Accuracy (pH):
±0.002

Accuracy (Salinity):
__

Alarm Output:
High/low limit alarm, calibration due alarm

Battery Life:
800 hr.

Calibration:
pH with calibration editing option, relative mV (RmV), ORP, dissolved oxygen and temperature

Calibration Points (Conductivity):
__

Cell Constant:
__

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Star A216 pH/RDO/DO benchtop meter with stand

Display Type:
LCD Graphic, Backlit

Measurement Modes:
pH, mV, Relative mV, ORP with Temperature, Dissolved Oxygen (% Saturation), Dissolved Oxygen (mg/L with Temperature)

Range (mV):
±2000.0 mV

Range (Resistivity):
__

Range (TDS):
__

Temperature Range (Metric):
-5.0°C to +105.0°C

Isopotential Point:
7.000pH

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Model:
Orion Star A216 pH/mV/RmV/ORP/RDO Optical/Dissolved Oxygen/Temperature Benchtop Meter

Range (ORP):
±2000.0 mV

Relative Humidity Range:
5 to 85 %, non-condensing

Resolution (mV):
0.1mV

Resolution (Salinity):
__

Salinity Measurement Modes:
__

TDS Factor Range:
__

Temperature Compensation:
__

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 Year Meter

Weight (English):
2 lb.

Outputs:
USB, RS232

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Resistivity):
__

Accuracy (TDS):
__

Backlight Option:
Yes, selectable

Battery Type:
4 AA (sold separately)

Calibration Points:
1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (pH):
1 to 5

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
2000 with date and time stamp

Channels:
2

IP Rating:
IP 54

Range (Conductivity):
__

Range (pH):
-2.000 to 20.000

Range (Salinity):
__

Resolution (pH):
0.1/0.01/0.001

Inputs:
BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), 9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature), stirrer input (stirrer probe)

Keypad:
Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m

Methods:
10 per channel with password protection

Power Supply:
Universal AC adapter (included) or 4 AA batteries (sold separately)

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
__

Resolution (Resistivity):
__

Resolution (TDS):
__

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
__

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
24cm x 18cm x 11cm (9.5 in. x 7.1 in. x 4.3 in.)

Weight (Metric):
0.9 kg

Img

Fischer Scientific

13-645-550

--


AC Adapter:
Universal 50-60 Hz, 100-240 VAC power adapter (Cat. No. 1010003)

Accuracy (Conductivity):
__

Accuracy (pH):
±0.002

Accuracy (Salinity):
__

Alarm Output:
High/low limit alarm, calibration due alarm

Battery Life:
800 hr.

Calibration:
pH with calibration editing option, relative mV (RmV), ORP, dissolved oxygen and temperature

Calibration Points (Conductivity):
__

Cell Constant:
__

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Star A216 pH/RDO/DO benchtop meter kit with ROSS Ultra Triode epoxy-body pH/ATC electrode, polarographic DO sensor, solutions and stand

Display Type:
LCD Graphic, Backlit

Measurement Modes:
pH, mV, Relative mV, ORP with Temperature, Dissolved Oxygen (% Saturation), Dissolved Oxygen (mg/L with Temperature)

Range (mV):
±2000.0 mV

Range (Resistivity):
__

Range (TDS):
__

Temperature Range (Metric):
-5.0°C to +105.0°C

Isopotential Point:
7.000pH

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Model:
Orion Star A216 pH/mV/RmV/ORP/RDO Optical/Dissolved Oxygen/Temperature Benchtop Meter

Range (ORP):
±2000.0 mV

Relative Humidity Range:
5 to 85 %, non-condensing

Resolution (mV):
0.1mV

Resolution (Salinity):
__

Salinity Measurement Modes:
__

TDS Factor Range:
__

Temperature Compensation:
__

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 Year Meter

Weight (English):
2 lb.

Outputs:
USB, RS232

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Resistivity):
__

Accuracy (TDS):
__

Backlight Option:
Yes, selectable

Battery Type:
4 AA (sold separately)

Calibration Points:
1 point relative mV (RmV) mode, 1 point ORP mode

Calibration Points (pH):
1 to 5

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
2000 with date and time stamp

Channels:
2

IP Rating:
IP 54

Range (Conductivity):
__

Range (pH):
-2.000 to 20.000

Range (Salinity):
__

Resolution (pH):
0.1/0.01/0.001

Inputs:
BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe), 9 pin MiniDIN (RDO optical or polarographic DO sensor with built-in temperature), stirrer input (stirrer probe)

Keypad:
Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m

Methods:
10 per channel with password protection

Power Supply:
Universal AC adapter (included) or 4 AA batteries (sold separately)

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
__

Resolution (Resistivity):
__

Resolution (TDS):
__

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
__

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
24cm x 18cm x 11cm (9.5 in. x 7.1 in. x 4.3 in.)

Weight (Metric):
0.9 kg

Img

Fischer Scientific

13-645-551

--


AC Adapter:
Universal 50-60 Hz, 100-240 VAC power adapter (sold separately, Cat. No. 1010003)

Accuracy (Conductivity):
__

Accuracy (pH):
±0.002

Accuracy (Salinity):
__

Alarm Output:
High/low limit alarm, calibration due alarm

Battery Life:
800 hr.

Calibration:
pH with calibration editing option, relative mV (RmV), ORP and temperature

Calibration Points (Conductivity):
__

Cell Constant:
__

Data Management Details:
Transfer single, range or all data points to printer or computer

Description:
Orion Star A321 pH portable meter

Display Type:
LCD Graphic, Backlit

Measurement Modes:
pH, mV, Relative mV, ORP with Temperature

Range (mV):
±2000.0 mV

Range (Resistivity):
__

Range (TDS):
__

Temperature Range (Metric):
-5.0°C to +105.0°C

Isopotential Point:
7.000pH

Memory:
Non-volatile memory preserves data log, calibration log and meter settings

Model:
Orion Star A321 pH/mV/RmV/ORP/Temperature Portable Meter

Range (ORP):
±2000.0 mV

Relative Humidity Range:
5 to 85%, non-condensing

Resolution (mV):
0.1mV

Resolution (Salinity):
__

Salinity Measurement Modes:
__

TDS Factor Range:
__

Temperature Compensation:
__

Temperature (Metric) Operating:
5°C to 45°C (ambient)

Temperature Probe Calibration:
1 point temperature offset calibration

Temperature Resolution:
0.1°C, 0.1°F

Warranty:
3 Year Meter

Weight (English):
1.2 lb.

Outputs:
USB, RS-232

Accuracy (mV):
±0.2 mV or ±0.05% of reading whichever is greater

Accuracy (Resistivity):
__

Accuracy (TDS):
__

Backlight Option:
Yes, selectable

Battery Type:
4 x AA

Calibration Points:
1 to 5

Calibration Points (pH):
__

Certifications/Compliance:
CE, TUV 3-1, FCC Class A

Data Points:
5000 with date and time stamp

Channels:
1

IP Rating:
IP 67

Range (Conductivity):
__

Range (pH):
-2 to 20

Range (Salinity):
__

Resolution (pH):
0.1/0.01/0.001

Inputs:
BNC (pH or ORP electrode), pin-tip (reference electrode), 8 pin MiniDIN (ATC temperature probe)

Keypad:
Comprehensive with menu-specific function keys and dual purpose scroll/shortcut keys

Log Function Type:
Automatic data logging with Auto-Read and Timed measure modes; manual data logging with Continuous m

Methods:
10 per channel with password protection

Power Supply:
4 AA batteries (included) or universal AC adapter (sold separately)

Range (Relative mV):
±2000.0 mV

Resolution (Conductivity):
__

Resolution (Resistivity):
__

Resolution (TDS):
__

Stability:
Selectable fast, medium, slow or Smart Stability setting; Selectable off or Automatic Smart averaging setting

Temperature Accuracy:
±0.1°C

Temperature (English) Operating:
41°F to 113°F (ambient)

Temperature (Metric) Reference:
__

Temperature Range (English):
22.0°F to 221.0° F

Temperature Selection:
Manual or automatic with ATC temperature probe

Dimensions (L x W x H):
24cm x 10cm x 6.5cm (9.5 in. x 3.9 in. x 2.6 in.)

Weight (Metric):